Operation Manual

44 GMC-I Messtechnik GmbH
V/Hz, Ω, Temperature, and A/Hz Measurements
5.7.4 Alternating Current Measurement with Clip-On Current Sensor,
A AC and Hz (METRAHITT
ECH and METRAHITBASE only)
Transformer Output, Voltage/Current
When a clip-on current sensor is connected to the multimeter
(METRAHITT
ECH: V input, METRAHITBASE: V input), all current
displays appear with the correct value in accordance with the se-
lected transformation factor.
The only prerequisite is that the current
sensor is equipped with at least one of the below listed transformation
factors, and that the factor has been previously selected in the following
menu (
Cl
ip
0FF
), see also chapter 6.4.
Current Clip Setup Menu
1nFo
... SET rAtE ... CLIP
1 / 10/100/1000 / 0ff

Maximum allowable operating voltage is equal to the current
transformer’s nominal voltage. When reading the measured value,
additional error resulting from the clip-on current sensor must also
be taken into consideration.
(default setting:
METRAHITTECH: OFF, METRAHITBASE: 1:1000)
Trans. Factors
CL IP
Measuring Ranges DMM Clip Types
100 mV 1 V 10 V
1:1
1 mV / 1 mA
100.00 mA 1.0000 A 10.000 A WZ12C
1:10
1m V / 10 mA
1.0000 A 10.000 A 100.00 A
WZ12B, Z201A/B,
METRAFLEX
1:100
1m V / 100 mA
10.000 A 100.00 A 1000.0 A
Z202A/B,
METRAFLEX
1:1000
1 mV / 1 A
100.00 A 1000.0 A (10000.0 A)
Z202A/B, Z203A/B
,
WZ12C,
METRAFLEX
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
A~
Hz
003.50
A
AC
TRMS
A~
Hz
1:1000
050.10
Hz
AC
1:1000
FUNC
ENTER
COM
V
Temp
Ω
V
~
R
x
Hz
Current
R
i
R
i
~ 1 M
Ω
METRAHITBASE
COM
V
Temp
Ω
A
~
R
x
Current
R
i
R
i
= 1 M
Ω/
9 M
Ω
METRAHITTECH
mA
METRAHITBASE
METRAHITTECH
V~
Hz
1MΩ
1kHz
V~
R
i
= 1 M
Ω
R
i
= 9 M
Ω