V/Hz, Ω, Temperature, and A/Hz Measurements
5.7.1 Direct and Pulsating Current Measurement, Direct Connection,
A DC and A (DC+AC)
(METRAHITX-
TRAOUTDOORTECHPRO only)
First disconnect supply power from the measuring circuit or
the power consumer (1), and discharge any capacitors.
In accordance with the current to be measured, turn the rotary
switch to A or A .
Select the current type appropriate for the measured quantity
by briefly pressing the FUNC | ENTER multifunction key. Each
time the key is pressed, the instrument is switched back and
forth between A DC and A (DC + AC)
TRMS
, which is indicated
by means of an acoustic signal. The current type is indicated at
the LCD by means of the DC or the (DC+AC)
TRMS
symbol.
Safely connect the measuring instrument (without contact
resistance) in series to the power consumer (2) as shown.
Switch supply power to the measuring circuit back on (3).
Read the display. Make a note of the measured value if the
instrument is not being operated in the memory mode or the
transmission mode.
Disconnect supply power from the measuring circuit or the
power consumer (1) once again, and discharge any
capacitors.
Remove the test probes from the measuring point and return
the measuring circuit to its normal condition.
COM mA A
V
Temp
Ω
~
R
x
~
R
x
1
3
2
I > 10 A
!
A
A
003
.
50
A
DC
003
.
50
A
DC
AC
TRMS
A
A
8
.
8
.
8
.
8
.
8
!
I > 16 A
Current
Measuring Ranges:
Current measurement
may only be performed
with installed batteries!
FUNC
ENTER
METRAHITX-TRA/OUTDOOR
METRAHITPRO
1 A / 10 A (16 A max. 30 s)
100 μA / 1 mA
10 mA / 100 mA
1 A / 10 A (16 A max. 30 s)
METRAHITTECH
10 mA / 100 mA
1 A / 10 A (16 A max. 30 s)