User Manual

42 GMC-I Gossen-Metrawatt GmbH
V/Hz, Ω, Temperature, and A/Hz Measurements
5.7.2 Alternating Current and Frequency Measurement,
Direct Connection, A AC and Hz
(METRAHITX-TRAOUTDOORTECHPRO only)
First disconnect supply power from the measuring circuit or
the power consumer (1), and discharge any capacitors.
In accordance with the current or frequency to be measured,
turn the rotary switch to A~ or Hz.
Select the desired measured quantity by briefly pressing the
FUNC | ENTER multifunction key. Each time the key is pressed,
AC
TRMS
and Hz are alternately selected, and
switching is acknowledged with an acoustic signal.
Safely connect the measuring instrument (without contact
resistance) in series to the power consumer as shown.
Switch supply power to the measuring circuit back on (3).
Read the display. Make a note of the measured value if the
instrument is not being operated in the memory mode or the
transmission mode.
Disconnect supply power from the measuring circuit or the
power consumer (1) once again, and discharge any
capacitors.
Remove the test probes from the measuring point and return
the measuring circuit to its normal condition.
COM mA A
V
Temp
Ω
~
R
x
~
R
x
1
3
2
A~
Hz
A~
Hz
003
.
50
A
AC
TRMS
050
.
10
Hz
AC
8
.
8
.
8
.
8
.
8
!
I > 10 A
!
I > 16 A
Current measurement
may only be performed
with installed batteries!
Measuring Ranges:
Current
FUNC
ENTER
METRAHITX-TRA/OUTDOOR
METRAHITPRO
1 A / 10 A (16 A max. 30 s)
100 μA / 1 mA
10 mA / 100 mA
1 A / 10 A (16 A max. 30 s)
METRAHITT
ECH
10 mA / 100 mA
1 A / 10 A (16 A max. 30 s)