Device and Measuring Parameters
A.diSP – analog display: select display mode
One of two different display modes can be selected for the
analog display:
bArG: bar graph
Point: pointer
1nFo
... SET rAtE ... A.diSP
bArG / Point

(bArG = default value)
APoFF – specified time for automatic shutdown and continuous ON
The instrument is switched off automatically if the measured value
remains unchanged for a long period of time, and if none of the
keys or the rotary switch have been activated before the specified
time “APoFF ” (entered in minutes) has elapsed.
If the on setting is selected, the multimeter is set to continuous
operation and
ON appears in the display to the right of the battery
symbol. In this case, the multimeter can only be switched off
manually. The “on ” setting can only be cancelled by changing the
respective parameter, and not by switching the instrument off.
1nFo
... SET rAtE ... APoFF
10 ... 59 min on

(10 minutes = default setting)
bEEP – set the limit value for continuity testing
1nFo
... SET rAtE ... bEEP
1, 10, 20, 30, 40, 90 Ω 
(10 Ω = default setting)
irStb – status of the infrared receiver in the stand-by mode
(METRAHITX-
TRAOUTDOOR only)
See chapter 7.2 on page 53 regarding settings.
Addr – select device address (METRAHITX-TRAOUTDOOR only
See chapter 7.2 on page 53.