Operation Manual

GMC-I Messtechnik GmbH 13
6.1 Transient Overvoltages
The multimeter is protected against transient voltages of up to 4 kV with front
and half times of up to 1.2 and 50 s respectively. Due to the fact that powerful
overvoltages must be reckoned with during measurement, for example in power
systems, at transformers or motors, we recommend the use of our KS30 mea-
suring adapter in such cases. It offers protection against transient overvoltages
of up to 6 kV with front and half times of up to 10 and 1000 s respectively.
Continuous load capacity is equal to 1200 V
eff
. Additional measuring error
due to use of the KS30 measuring adapter amounts to approximately –2%.
6.2 Measuring Voltages of Greater than 300 V
Voltages of greater than 300 V can be measured with a high-voltage measuring
probe, for example the HV3
1)
of the HV30
2)
from GMC-I Messtechnik GmbH.
The bonding terminal must be grounded in this case. Observe all applicable
safety precautions!
1)
HV3: 3 kV
2)
HV30: 30 kV for (DC) voltage only
6.3 Low-Voltage Measurement
The instrument is equipped with a special 30 mV measuring range for measuring
voltage drop at fuses which is distinguished by high resolution (10 V) with a low
input resistance of 50 k.
Set the rotary switch to “Temp RTD”.
Select “V DC” measurement with probe by repeatedly pressing the FUNC
key until “mV DC” appears at the display.
Connect
the probe to the ” and “V” sockets.
7 Current Measurement
First disconnect supply power from the measuring circuit or the power
consumer and discharge all capacitors, if any are present.
Depending upon the type of current, set the rotary switch toA ” or “A .
The symbol for the selected current type, (DC) or ~ (AC), appears at the LCD.
Connect the measuring instrument securely to the power consumer in se-
ries as shown in the diagram (without transition resistor).
black
black
red
x1000
x100
Measuring voltages of greater than 300 V
with the HV3 high-voltage probe