Datasheet

METRISO 1000A
Insulation Tester
2 GMC-I Messtechnik GmbH
Technical Data
1)
)
As related to scale length
Scale lengths: R1 l = 46 mm
R2 l = 71 mm
R3 l = 80 mm
l = 67 mm
Ul = 66 mm
2)
Within the marked area of the respective scale (nominal range of use)
Accuracy
Insulation resistance Class 1.5 in reference to scale length
after electrical zero balance
Direct and
alternating current class 2.5
Display
Movement Moving coil with core magnet
Turn-on performance
for insulation resistance measurement
An electronic voltage limiter prevents the voltage at the device
under test from substantially exceeding nominal voltage at turn-
on (see following characteristic curves).
Measuring
Function
Measuring Range
Intrinsic
Uncertainty at
Reference
Conditions
1)
Measuring
Uncer-
tainty
2)
Nominal
Voltage
U
N
Nominal/
Measuring
Current
Open-
circuit
Voltage
U
o
Fre-
quency/
Short-
circuit
Current I
k
Internal
Resistance
R
i
Switching Value
for Resistance
Overload Capacity
Value Duration
1000 V 0 ... 1000 V 2.5 %
DC /
40 ... 200
Hz
900 k 1200 V
continu-
ous
R1, R2, R3
U
N
= 50 V
0 ... 40 k
1.5 %
25%
rdg.
50 V
I
N
1.0 mA
60 V
<12mA
30 k
> 100 k 1200 V max. 10 s20 k ... 1 M 10 k
200 k ... 20 M 40 k
R1, R2, R3
U
N
= 100 V
0 ... 80 k
1.5 %
25%
rdg.
100 V 120 V
60 k
> 200 k 1200 V max. 10 s40 k ... 2 M 20 k
400 k ... 40 M 80 k
R1, R2, R3
U
N
= 250 V
0 ... 200 k
1.5 %
25%
rdg.
250 V 300 V
150 k
> 500 k 1200 V max. 10 s100 k ... 5 M 50 k
1 M... 100 M 200 k
R1, R2, R3
U
N
= 500 V
0 ... 400 k
1.5 %
25%
rdg.
500 V 600 V
300 k
> 1 M 1200 V max. 10 s200 k ...10 M 100 k
2 M ... 200 M 400 k
R1, R2, R3
U
N
= 1000 V
0 ... 0.8 M
1.5 %
25%
rdg.
1000 V 1200 V
600 k
> 2 M 1200 V max. 10 s400 k ... 20 M 200 k
4 M ... 400 M 800 k
4 0 ... 4  1.5 %
10%
rdg.
—I
m
200 mA 9 V > 200 mA 0,315 A
continu-
ous
Ux
Rx
[V]
500
0
U
N
= 500 V
400
300
200
100
500
1000
Range R1
Ux
Rx
[V]
500
0
U
N
= 500 V
400
300
200
100
Range R3
600
Ux
Rx
[V]
500
0
U
N
= 500 V
400
300
200
100
0,5
1
Range R2
600
1,5
2
[M]
[k]
[M]