Datasheet

GMC-I Messtechnik GmbH 5
METRISO PRIME+
Digital High-Voltage Insulation Tester
Features and Accessories for High-Voltage Tests
Features B1 (PROFITEST 204HP-2.5kV) and B2 (204HV-5.4kV)
Expanded features for high-voltage tests
Test voltage selectable in 50 V steps
Rise time (ramp) adjustable from 0.1 to 99 s
Test duration adjustable from 1 to 120 s
Floating test voltage outputs
Electronically controlled test sequence
Test sequence can be started with test pistol
Breakdown voltage display
Pulse-arc operation
Phase angle display
Measured values can be saved to memory
Acoustic and optical error messages
Key switch for protection against unauthorized start-up
Connector terminals for external signal lamps
Features B1 (PROFITEST 204HP-2.5kV)
Voltage test per EN 60204 / VDE 0113
Test power: 500 VA (intermittent)
Breaking current adjustable in 1 mA steps
Features B2 (PROFITEST 204HV-5.4kV)
Test power: 50 VA
Breaking current adjustable in 0.5 mA steps
The respective high-voltage module which is supplied ex works
firmly attached to the MERISO PRIME+ basic instrument (i.e. it
cannot be removed or subsequently fitted be the user) is intended
for high-voltage testing.
Voltage, current and phase angle can be measured via fixed mea-
suring cables.
Technical Data PROFITEST 204HP-2.5kV
Technical Data PROFITEST 204HV-5.4 kV
Feature D: Trolley Caddy 204
Trolley for basic instrument
combined with high-voltage
module, includes cover with
side pockets.
Feature F1: Signal 204
Signal lamp set mounted
to a magnetic base for
indicating high-voltage
testing in accordance with
DIN VDE 0104.
Accessory: Claim 204
Set of various items used
to warn unauthorized per-
sons and for securing
large areas, machines or
machine components
during the performance of
high-voltage testing.
Nominal Range
of Use
Resolu-
tion
Measuring
Uncertainty
Intrinsic
Uncertainty
Test Vol tage
U AC
250 V
2.5 kV
1V
10 V
±(5% rdg. + 5 d)
±
(2.5% rdg. + 5 d)
Meas.
Quantity
Current
I AC
10.0 200 mA
0.1 mA
1mA
±(7% rdg.+ 5 d) ±(5% rdg. + 5 d)
Nominal Range
of Use
Resolu-
tion
Measuring
Uncertainty
Intrinsic
Uncertainty
Test Vol tage
U AC
650 V1.00 kV
1.00 kV5.35 kV
1V
10 V
+2 7% rdg.
+2 5% rdg.
0 5% rdg.
0 3% rdg.
Meas.
Quantity
Current
I AC
1.0 10.0 mA
0.01 mA
0.1 mA