Datasheet

METRISO PRO
Analog Insulation, Low Resistance and Voltage Measurement Instrument
2 GMC-I Messtechnik GmbH
Characteristic Values
1
Accuracy specified here is only achieved with the following optional accessory:
“shielded high-resistance measurement cable KS-C (article number Z541F)”.
2
Relative to scale length:
Scale 1: 83.13 mm
Scale 2: 75.05 mm
Scales 3 and 4: 67.02 mm
3
Within the identified range on the respective scale (nominal range of use)
4
with ZERO balancing
5
0.17 ... 220%
Displays
Analog Display
Measuring movement Moving-coil mechanism with core
magnet
Scale length 83.13 mm (longest scale)
Limit LED LED lights up red to indicate an
exceeded limit value
LED lights up green to indicate adher-
ence to the limit value
LED LED lights up red to indicate the pres-
ence of interference voltage (when
instrument is switched off),
test voltage during insulation measure-
ment or residual voltage after insulation
testing (U > 50 V),
Reference Conditions
Reference temperature + 23 C 3K
Relative atmospheric
humidity 40 ... 75%
Measured quantity
frequency
45 Hz ... 65 Hz
Measured quantity
waveform Sine, deviation between TRMS and
rectified value < 1%
Battery voltage 9.5 V 0.1 V
Test resistance 10 M1%
Normal position of use 30°
Power Supply
Batteries 8 ea. 1.5 V mignon cell (8 ea. size AA)
(alkaline manganese per IEC LR14) or 8
rechargeable NiMH batteries (must be
recharged externally)
Nominal range of use
8.5 ... 12 V
Battery test Battery capacity display via the
Bat. TEST key
Battery saver circuit
Standby function
Service life For R
INS
(1000 V / 1 M) and R
LO
with
20 s on-time and 1 measurement each
for a duration of 5 s
– With batteries (alkaline manganese):
900 measurements
– With rechargeable batteries
(2000 mAh):
850 measurements
Safety shutdown If supply voltage is too low, the instru-
ment is switched off, or cannot be
switched on.
When the rotary switch is set to the OFF
position, the instrument is completely
disconnected from the batteries (after
approximately 10 seconds).
Meas.
Qty.
Scale /
Standard
Measuring
Range
Nominal range of
use
Nominal / Open-
Circuit Voltage
Nominal
Current I
N
Short-
Circuit
Current I
K
Intrinsic Uncertainty
under Reference
Conditions
2
Measuring
Uncertainty
3
Overload
Capacity
R
INS
VDE 0413
100 k ...
100 M
100 k ...
10 M
50 V /100 V:
1.25 U
ISO
250 V /500 V /
1000 V:
1.1 U
ISO
1 mA 5 mA
2.5%
30% of the
measured value
1000 V AC/DC
TRMS
10 k ... 1 T
100 k ...
100 G
2.5%
1
R
LO
0 to 5 0.17 to 4 
4
4 V < U
0
< 6 V
Test current I
N
200 mA 2.5%
10%
5
of the
measured value
1000 V AC/DC
TRMS
U
AC/DC
0 ... 1000 V 10 ... 1000 V 2.5%
1000 V AC/DC
TRMS
U
BAT
8 ... 12 V 8.5 ... 12 V
2
1
3
4
5
1
2
4
5
3
!