SECULIFE ESPRO
TYPICAL TEST SETUP
Unlike all conventional ESU Analyzers with lesser degrees of accuracy, the SECULIFE ESPRO utilizes
an external Current Transformer and external precision load resistors (values to be determined by the
manufacturer’s suggested test load for the generator you are testing or servicing) for typical
Electrosurgical Generator testing. Many of the world’s leading Electrosurgical generator manufacturers
utilize this exact same technique when they test, service and calibrate their generators.
This methodology has several distinct advantages over conventional ESU Analyzers:
Improved accuracy and resolution capabilities
100% manufacturer recommended test load compatibility
Smaller and lighter weight instrumentation
The current transformer senses the RF current flowing through the external test load and produces a
proportional voltage as an input to the
SECULIFE ESPRO. This input is either a direct 1 : 1 input (for
current transformers with a ratio of 1 : 1 volts : amps) or a 0.1 : 1 input (for current transformers with a
ratio of 0.1 : 1 volts : amps). Combining the standard and low ranges of the
SECULIFE ESPRO with
the use of either a 0.1:1 or 1:1 ratio current transformer, the user has full control over the ability to get
high accuracy and high resolution readings from all types of Electrosurgical Generators.