Operation Manual

SECULIFE ESPRO
Typical Setup Screen
The following is a breakdown of the parameters available in the configuration of the unit and their
available options:
Factory Setup Configuration
Parameter Description Range
Input Range
Used to select the input range that is to be
calibrated.
100 mV
1000 mV
Donut Attenuation
Selects the Attenuation in Volts : Amp for the RF
Current Transformer
being used. Each setting
has an independent gain setting.
0.1 : 1
1 : 1
Volts : Amps
Counts Zero Offset
Used to adjust the A/D converter output to zero
counts when zero input is applied. This should
be adjusted to get the smallest mVRMS reading
when the input is zero.
1000 Counts
Input Gain
Used to calibrate the mV reading for when the
0.1 : 1 RF Current Transformer
(Donut)
attenuation is selected.
0 to 5000
mV RMS Reading
Shows the latest measurement by the system.
This is for display only to help in calibrating the
system.
0.0 to 700.0 mV
Input Zero
Zeros the input circuitry based on donut being
used. Each donut can have a slightly different
zero offset. This function will eliminate the offset
from the readings. Press the SETUP key while
this parameter is selected to perform the auto –
zero function. This will also remove any offsets
created by the User in the System Setup screen.
Press Setup
28 GMC-I Messtechnik GmbH