Operation Manual
SECULIFE ESPRO
SYSTem Subsystem
This group allows the user to setup the startup mode for the unit, as well as directly control
the unit, as if pressing the keys on the front panel.
KEYWORD PARAMETER FORM COMMENTS
SYSTem:
POWer DEFaults | CUStom | SETCurrent
CONtrast < numeric_value > Numbers 1-20
KEY DUP | DDN | SUP | SDN | VUP |
VDN | SETup | RETurn | GSAVe |
HOLD
VER? Read only
READ Subsystem
This group allows the user to get measurements from the unit.
KEYWORD PARAMETER FORM COMMENTS
READ:
MVrms | MArms | Warms | MVPeak |
MVPP | MVP-PP | MVPK+ | CF |
TON | TOFF | TCYC | DCYC |
MVCyc | MACyc | WCyc Read only
DATA Read only
LOCn Read only, ānā = stored
waveform location to read (1-3)
STATus Sub-system
This subsystem provides status on the operating mode of the unit including messages that would
normally be seen on the display.
KEYWORD PARAMETER FORM COMMENTS
STATus? Read Only
Bit Value Definition
0 1 Hold Mode
1 2 Graph Mode
2 4 Calibration Mode
3 8
4 16
5 32
6 64
7 128
8 256 Error Present
9 512 mV Out of Range
10 1024 mA Out of Range
11 2048 Watts Out of Range
12 4096 mV Peak Out of Range
13 8192 Crest Factor Out of Range
14 16384
15 32768
GMC-I Messtechnik GmbH 43