Operation Manual
SECULIFE ESPRO
GMC-I Messtechnik GmbH 45
SECULIFE ESPRO Communication Command Summary
Keywords Nodes
Subnodes
Values
MVrms? Returns: mV RMS [read only]
MArms? Returns: mA RMS [read only]
WArms? Returns: Watts RMS [read only]
MVPeak? Returns: mV Peak [read only]
MVPP? Returns: mV Peak to Peak [read only]
MVP-PP? Returns: mV Peak/Peak to Peak [read only]
MVPK+? Returns: mV Positive Peak [read only]
CF? Returns: Crest Factor [read only]
TON? Returns: Time - Pulse On [read only]
TOFF? Returns: Time - Pulse Off [read only]
TCYC? Returns: Time - Total Cycle [read only]
DCYC? Returns: % Duty Cycle [read only]
MVCyc? Returns: mV Pulse Cycle [read only]
MACyc? Returns: mA Pulse Cycle [read only]
WCyc? Returns: Watts Pulse Cycle [read only]
DATA?
Returns: Entire Data Buffer [read only] Length = NUMSamples * 2
Data Format is mV in signed double byte format, Decimal place is assumed based on
Input Range (Low Range = 2DP, High Range = 1DP)
READ
LOCn?
Returns: Saved Data Buffer at Location 'n' , n= 1-3 [read only] Length = NUMSamples * 2
Data Format is mV in signed double byte format, Decimal place is assumed based on
Input Range (Low Range = 2DP, High Range = 1DP)
Bit Value Definition
0 1 Hold Mode
1 2 Graph Mode
2 4 Calibration Mode
3 8
4 16
5 32
6 64
7 128
8 256 Error Present
9 512 mV Out of Range
10 1024 mA Out of Range
11 2048 Watts Out of Range
12 4096 mV Peak Out of Range
13 8192 Crest Factor Out of Range
14 16384
STATus?
15 32768