Pressing on the caution icon will display the active faults and the status of the faults. When a fault has
been cleared, such as a blocked fan or system overtemperature, pressing the Reset Cleared Faults
button will remove the fault from the list. Once all faults are cleared, the caution icon will be removed from
the main screen.
The following table shows all of the faults that are monitored by the system.
Message
Description
Recovery Action
Left Fan Blocked
The left fan is not spinning. Permanent
damage or complete failure could occur.
Check the fan for obstructions. If nothing is
found and the fan is not spinning contact
GMC-I Service GmbH for service.
Right Fan Blocked
The right fan is not spinning. Permanent
damage or complete failure could occur.
Check the fan for obstructions. If nothing is
found and the fan is not spinning contact
GMC-I Service GmbH for service.
System Overtemp
The internal temperature has exceeded
normal operating conditions.
Stop activating the DUT, allow the system to
cool for at least a half hour. If the problem
persists, contact GMC-I Service GmbH for
service.
Footswitch Fuse
Open
The internal fuse protecting the power supply
to the footswitch interface has opened.
Contact GMC-I Service GmbH for service.
Front USB Fuse
Open
The internal fuse protecting the power supply
to the front USB port has opened.
Contact GMC-I Service GmbH for service.
Side USB Fuse
(Front) Open
The internal fuse protecting the power supply
to the side USB port towards the front of the
unit has opened.
Contact GMC-I Service GmbH for service.
Side USB Fuse
(Rear) Open
The internal fuse protecting the power supply
to the side USB port towards the rear of the
unit has opened.
Contact GMC-I Service GmbH for service.
PS/2 Fuse Open
The internal fuse protecting the power supply
to the PS/2 port has opened.
Contact GMC-I Service GmbH for service.