Operation Manual

108 GMC-I Messtechnik GmbH
SECULIFEESXTRA
SERIAL COMMUNICATION - There is a serial port on the side panel. The RS-232 Port is used to
communicate with the DUT.
USB COMMUNICATION - There are two USB ports on the side panel and one on the front. The USB
Ports are used to connect devices such as keyboards, mice, PCL printers, or flash drives.
ETHERNET COMMUNICATION - There is a 10/100 Ethernet port on the side panel. This is used for
firmware updates or for connection to a LAN for data storage, remote control, and printing test records.
POWER SWITCH - The main power switch for the Analyzer is located on the front panel. When the
switch is ON, the LED should be ON.
LINE POWER - A Kycon 3 position locking receptacle is provided for the 12 VDC Universal Power Supply
input. Two receptacles are available, one on the side of the unit and one on the rear.
NOTE: ONLY ONE POWER RECEPTACLE SHOULD BE USED AT ANY TIME.
No Response
from DUT
The SECULIFE ESXTRA expected RS-232
data from the DUT and did not receive it.
Check the RS-232 cable to ensure that
it is seated completely on the
SECULIFE ES
XTRA and DUT.
ForceTriad
Version Error
The Autosequence is only compatible with
ForceTriads running firmware 3.50.
This error will occur if you are trying to use the
ForceTriad PM Autosequence on a unit with
software version less than 3.50.
Go to www.valleylab.com to get a software
update for your ForceTriad.
SYSTEM INPUTS AND OUTPUTS
Message
Description
Recovery Action