CONFigure Subsystem
This group allows the user to setup the display and operational settings for the unit. Note that there are
independent settings for the Measure RF Energy screen and the Measure RF Leakage screen.
KEYWORD PARAMETER FORM COMMENTS
CONFigure
:RFMeasure
:DISPlay :SxZy nn S = display_screen_number
Z = zone_number
:SCReen < numeric_value > Range 1-6
1-5 = # display zones
6=Measurement List Display
:AVERaging FAST | SLOW | MEDium
:LOAD :MODE INTernal | EXTernal | INT/EXTernal
:INTernal <numeric_value> 0-6400
:EXTernal <numeric_value> 0-6400
:ACTual? Read Only, returns calibrated
load value, what the
SECULIFE ES
XTRA uses
for Watts calculations
INPut :RANGe 100 | 1000 | AUTo
:MODE CONTinuous | PULsed
:TYPE ISOlated | GND
display_screen_number = 1-6
1 = One Parameter
2 = Two Parameters
3 = Three Parameters
4 = Four Parameters
5 = Five Parameters
6 = Display All Parameters
nn = Parameter for selected Zone
0 = mV RMS 8 = kHz
1 = mA RMS 9 = Time-Pulse On
2 = Watts RMS 10 = Time-Pulse Off
3 = mV Peak 11 = Time-Total Cycle
4 = mV Pk-Pk 12 = % Duty Cycle
5 = mV Pk / mV Pk-Pk 13 = mV Pulse Cycle
6 = mV Pk+ 14 = mA Pulse Cycle
7 = Crest Factor 15 = Watts Pulse Cycle