Operation Manual
GMC-I Messtechnik GmbH 115
SECULIFEESXTRA
CONFigure
:RFLeakage
:MODE LKG1A | LKG1B | LKG2 | LKG3
:DISPlay :SxZy nn S = display_screen_number
Z = zone_number
:SCReen < numeric_value > Range 1-6
1-5 = # display zones
6=Measurement List Display
:AVERaging FAST | SLOW | MEDium
:LOAD :MODE INTernal | EXTernal | INT/EXTernal
:INTernal <numeric_value> 0-6400
:EXTernal <numeric_value> 0-6400
:TOTal? Read Only, returns combined
load based on internal
INPut :RANGe 100 | 1000 | AUTo
:MODE CONTinuous | PULsed
display_screen_number = 1-6
1 = One Parameter
2 = Two Parameters
3 = Three Parameters
4 = Four Parameters
5 = Five Parameters
6 = Display All Parameters
nn = Parameter for selected Zone
0 = mV RMS 8 = kHz
1 = mA RMS 9 = Time-Pulse On
2 = Watts RMS 10 = Time-Pulse Off
3 = mV Peak 11 = Time-Total Cycle
4 = mV Pk-Pk 12 = % Duty Cycle
5 = mV Pk / mV Pk-Pk 13 = mV Pulse Cycle
6 = mV Pk+ 14 = mA Pulse Cycle
7 = Crest Factor 15 = Watts Pulse Cycle