Operation Manual
GMC-I Messtechnik GmbH 119
SECULIFEESXTRA
Common Command Group
This subsystem provides access to common SCPI commands.
KEYWORD COMMENTS
*IDN? Read Only, Returns the following information:
Manufacturer, Model, Serial Number, Firmware Version, Software Version, OS Version.
*RST Write Only, Resets unit to default power up state.
If selected, the saved profile is loaded, see Profiles on page 90.
*STB? Read Only, Returns the status byte information:
Bit Value Definition
0 1 Hold Mode
12
2 4 Calibration Mode
38
4 16 Tissue Response Mode
532
6 64 Tissue Response Test Complete
7 128 Tissue Response Test Started
8 256 Error Present
9 512 mV Out of Range
10 1024 mA Out of Range
11 2048 Watts Out of Range
12 4096 mV Peak Out of Range
13 8192 Crest Factor Out of Range
14 16384 Fan 1 Blocked
15 32768 Fan 2 Blocked