Operation Manual
12 GMC-I Messtechnik GmbH
SECULIFEESXTRA
OPTIONAL ACCESSORIES (continued):
Footswitch Simulator Tissue Response Load
CQM Cable Differences - Two CQM cables are provided in the standard accessory kit. The only
difference between them is that one cable has a small pin and the other does not. The cable with the pin
is used to enable the REM circuitry on the DUT. The cable without the pin is used to bypass the REM
circuit and allow for shorting of the REM leads. Note that some generators do not allow for the REM
circuit to be disabled. For these generators it does not matter which cable is used.
When using the SECULIFE ES
XTRA CQM test mode, the cable with the pin must be used. For other
modes, the Dispersive ports on the SECULIFE ES
XTRA are shorted together and the cable without the pin
should be used.
Footswitch
Simulator
Tissue Response
Load
Cable Without Pin,
Disables REM
Cable With Pin,
Enables REM