Operation Manual
124 GMC-I Messtechnik GmbH
SECULIFEESXTRA
The RF Measure screen shows mA, but I
need to measure watts. Can the
SECULIFE ES
XTRA measure watts?
Yes, press the mA RMS text to select the measurement to be shown.
You can also press the Display Parameters button to select how
many measurements are shown on the screen at a time.
Why do I have two blue REM cables?
The cable with the clear pin is used to enable the dual pad function
(REM) on the DUT. This cable is used for REM test mode only. For
RF Power measurements the SECULIFE ES
XTRA shorts the two
dispersive inputs internally. The blue cable without the pin is used
because it disables the REM function on the DUT.
I am testing a ForceTriad. When I key the
generator, I get a REM alarm on the
ForceTriad.
You must configure the DUT for Demo mode to bypass the REM
alarm. Use the Tool icon on the right most display and select Demo
Mode from the menu on the left most display.
The SECULIFE ES
XTRA shows a mV
RMS measurement. Is this the load
voltage?
No, this is the output voltage of the internal Current to Voltage RF
transformer. This was included for compatibility with the SECULIFE
ESPRO analyzer. To get load voltage, multiply the RMS current by
the load resistance.
Question Answer