Operation Manual

126 GMC-I Messtechnik GmbH
SECULIFEESXTRA
SPECIFICATIONS
RF MEASUREMENT
INPUT
IMPEDANCE
0-6400
CONNECTION 4mm Safety Jack
MAXIMUM
VOLTAGE
10kV
FREQUENCY 10 kHz - 10 MHz
MEASUREMENT
METHOD
Pearson current to voltage converter
0.1V : 1A
VOLTAGE (RMS)
100 mV RANGE
0.20 - 70.00 mV RMS
0.01 mV Resolution
1000 mV RANGE
2.0 - 700.0 mV RMS
0.1 mV Resolution
VOLTAGE
(pk, pk-pk)
100 mV RANGE
0.20 - 100.0 mV
0.01 mV Resolution
1000 mV RANGE
2.0 - 1000.0 mV
0.1 mV Resolution
ACCURACY,
100 mV RANGE
f 1.0 MHz
1.0 MHz < f
2.5 MHz
f > 2.5 MHz
Input
10.0 mV
± 1% reading or
±0.25 mV
± 1 % reading or
± 0.25 mV
± 1 % reading or
± 1.0 mV
10.0 mV < Input
35.0 mV
± 1 % reading
or ± 0.25 mV
± (1% reading
+ 0.5 % range)
± (6 % reading +
0.5 % range)
Input
> 35.0 mV
± 1 % reading or
±0.25 mV
± (1 % reading +
0.5 % range)
± (6 % reading + 5 %
range)
ACCURACY,
1000 mV RANGE
ƒ 2.5 MHz ƒ 2.5 MHz
Input 50 mV
± 1% reading
or ± 1 mV
± 1% reading
or ± 1 mV
50 mV < Input
400.0 mV
± 1% reading ± 4% reading
Input > 400.0 mV ± (1% reading + 0.25% range)
± (4% reading +
0.25% range)