PHYSICAL, ENVIRONMENTAL AND ELECTRICAL
CONSTRUCTION Enclosure - aluminum
OPERATING RANGE
15 - 30 °C
20 - 80% RH, Non-Condensing
STORAGE RANGE -20 - 60 °C
SIZE 198.1 x 381 x 571.5 mm
CONSTRUCTION
ENCLOSURE ABS Plastic
FACE Lexan, Back Printed
WEIGHT 14 kg
POWER
12 VDC, minimum 4A
Kycon 3 position locking receptacle
ESU ANALYZER MODES
RF MEASUREMENT MODE
Allows use as a general purpose RF meter, including current, voltage,
power and timing measurement.
LEAKAGE TEST 1A
Tests the open circuit leakage of an Isolated Type CF generator. The test
complies with IEC 601.2.2, sec. 19.101b, fig, 104 and sec. 19.102,
adopted by ANSI/AAMI HF18-2001.
LEAKAGE TEST 1B
Tests the open circuit leakage of an Isolated Type CF generator. The test
complies with IEC 601.2.2, sec. 19.101b, fig, 104 and sec. 19.102,
adopted by ANSI/AAMI HF18-2001.
LEAKAGE TEST 2
Tests the leakage to ground of Earth-referenced Type BF generators. This
test complies with IEC 601.2.2, sec. 19.101a, test 1, fig. 102, and sec.
19.102, adopted by ANSI/AAMI HF18-2001.
LEAKAGE TEST 3
Tests the leakage to ground of Earth-Referenced Type BF generators.
This test complies with IEC 601.2.2, sec. 19.101a, test 2, fig. 102, and sec.
19.102, adopted by ANSI/AAMI HF18-2001.
LOAD CURVE
Allows user to automatically generate a power curve based on varying
load, commonly seen in the generator data sheets.
REM/ARM/CQM MODE
Uses internal load bank as a variable resistor to test the Generator's
CQM/RECQM circuit.
AUTOSEQUENCE MODE
Allows user to follow a standard or custom defined protocol to sequentially
buildup a test record, applicable to many manufacturer's recommended
verification and calibration procedures.