Operation Manual
14 GMC-I Messtechnik GmbH
SECULIFEESXTRA
The SECULIFE ESXTRA utilizes an internal Current Transformer and internal precision load resistors for
simple configuration of typical Electrosurgical Generator testing. Many of the world's leading
Electrosurgical generator manufacturers utilize this exact same technique when they test, service and
calibrate their generators.
Internal configuration relays allow for simple connections to the DUT. When the Measure RF Energy,
Measure RF Leakage, or REM/ARM/CQM mode is selected from the main menu, the relays configure the
internal connections as required. Shown below are examples of external connections to the DUT.
Monopolar:
1) Yellow Active Lead from Active Jack on SECULIFE ES
XTRA to DUT Output.
2) Blue Dispersive Lead from Dispersive Jack on SECULIFE ES
XTRA to DUT Dispersive.
TYPICAL MEASUREMENT CONNECTIONS
ESU (DUT)
Patient Return/
Dispersive
Electrode
ESU (DUT)
Active
Electrode