Tissue Test - This mode evaluates the tissue response function of the DUT. In this mode, the SECULIFE
ES
XTRA is configured for an external TRL-2420 lamp load. The SECULIFE ESXTRA is connected in series
with the load as shown below. The terminals are labeled Return and Active for reference only. The DUT
Return terminal shown below is typically not the Patient Return or REM electrode. Refer to the DUT
service manual to determine which DUT ports are used for the tissue response output.
If the output terminals of the DUT are not compatible with the standard SECULIFE ES
XTRA safety test
leads, the alligator clips supplied with the SECULIFE ES
XTRA accessory kit can be used to connect
directly to the electrosurgical instrument as shown in the connection diagram below.
ESU Generator
(DUT)
Return Active