Operation Manual
22 GMC-I Messtechnik GmbH
SECULIFEESXTRA
Leakage 3 - This leakage test, specified by the IEC as Earth Reference Leakage Type BF (Load from
Active Electrode to Earth) is for testing the leakage to earth ground of a Ground Referenced Output type
BF electrosurgical generator from the active output. This test complies with IEC 601.2.2, sec. 19.101a,
test 2, fig. 102, and sec. 19.102, adopted by ANSI/AAMI HF18-2001. The purpose of this test is to verify
that the RF leakage of the Device Under Test (DUT) meets or exceeds the IEC specification.
Connections
1) Active Lead from Active Jack on SECULIFE ESXTRA to DUT Active Output.
2) Dispersive Lead from Dispersive Jack on SECULIFE ES
XTRA to DUT Dispersive.
ESU (DUT)
Patient Return /
Dispersive
Electrode
ESU (DUT)
Active
Electrode