Operation Manual

GMC-I Messtechnik GmbH 23
SECULIFEESXTRA
This section was written to help you get started with the SECULIFE ESXTRA. Examples are given for
typical configurations of RF Measurement, REM/ARM/CQM testing, Load Curves and Autosequences.
This section will give specific examples of how to configure the SECULIFE ES
XTRA plus additional
modifications that might help fulfill your specific testing requirements.
Scenario 1 - I need to measure monopolar current with a 500 ohm load. I don't have a footswitch cable,
so I will trigger the DUT myself.
Solution:
1) Select Measure RF Energy from the main menu.
2) Connect the Active output from the DUT to the Active input on the SECULIFE ES
XTRA. If you are
using the cables from the SECULIFE ES
XTRA accessory kit, use the Yellow Lead.
3) Connect the Dispersive or Return port on the DUT to the Dispersive 1 and Dispersive 2 inputs on the
SECULIFE ES
XTRA. If you are using the cables from the SECULIFE ESXTRA accessory kit, use the
Blue cable without the pin.
4) Press the Isolated Generator Type button to configure the SECULIFE ES
XTRA for isolated generator
measurements.
NOTE: To see how the SECULIFE ES
XTRA is internally configured, press the Test Setup button. To close
the image, press Test Setup again.
GETTING STARTED
MEASURE RF ENERGY