Operation Manual
GMC-I Messtechnik GmbH 25
SECULIFEESXTRA
Scenario 2 - I need to measure mA and Watts of a Bipolar output with a 300 ohm Load. I want the
SECULIFE ES
XTRA to trigger the DUT with the footswitch output.
Solution:
1) Select Measure RF Energy from the main menu.
2) Connect one Bipolar output from the DUT to the Active input on the SECULIFE ES
XTRA. If you are
using the cables from the SECULIFE ES
XTRA accessory kit, use the Yellow Lead.
3) Connect the other Bipolar output port from the DUT to the Dispersive 1 input on the SECULIFE
ES
XTRA. If you are using the cables from the SECULIFE ESXTRA accessory kit, use the Blue cable
with banana jacks on each end, not the CQM/RECQM cable .
4) Press the Test Load button to open the load configuration box. The screen should look like the
following:
5) Press the Button next to Internal Load and enter 300 in the number pad. The load configuration box
should automatically close.