Operation Manual
GMC-I Messtechnik GmbH 27
SECULIFEESXTRA
7) Press the Configure Footswitch button to open the Footswitch Setup box.
8) Use the dropdown arrow under Select Output to select the Biploar output.
9) Use the dropdown arrow under Activation Timeout to select the maximum time the DUT will be
triggered.
NOTE: The footswitch output selection is only valid for footswitch cables that are designed for a specific
DUT, such as the Z699B or Z699C. If you are using the unterminated footswitch cable, Z699A, the output
selection is dependent on how the cable is wired.
10)Press the footswitch button to Key the DUT. The mA and Watts measurements will be shown on the
SECULIFE ES
XTRA Screen.
11)You may either wait for the Footswitch Activation Timeout, or press the Footswitch button again to
stop keying the DUT output.