Scenario 3 – I need to measure the mA output of a Hyfrecator (ground referenced generator) with a 100
ohm load. I don’t have a footswitch cable, so I will trigger the DUT myself.
Solution:
1) Select Measure RF Energy from the main menu.
2) Connect the Active output from the DUT to the Active input on the SECULIFE ES
XTRA. If you are
using the cables from the SECULIFE ES
XTRA accessory kit, use the Yellow Lead.
3) Press the Gnd Ref generator type button to configure the SECULIFE ES
XTRA to make a ground
referenced measurement.
NOTE: To see how the SECULIFE ES
XTRA is internally configured, press the Test Setup button. To close
the image, press Test Setup again.