Leakage Test 1a - I need to measure the leakage current from an Active lead to ground of a type CF ESU
generator.
Solution:
1) Select Measure RF Leakage from the main menu.
2) If the Leakage selection at the top of the screen does not show Leakage 1a, press the button and
select Leakage 1a from the test mode menu.
3) Connect the Active output from the DUT to the Active input on the SECULIFE ES
XTRA. If you are
using the cables from the SECULIFE ES
XTRA accessory kit, use the Yellow Lead.
4) Connect earth ground port on the SECULIFE ES
XTRA to the ground lug on the back of the DUT. If you
are using the cables from the SECULIFE ES
XTRA accessory kit, use the green cable and green
alligator clip.
NOTE: To see how the SECULIFE ES
XTRA is internally configured, press the Test Setup button. To close
the image, press Test Setup again.
5) Key the DUT. The SECULIFE ES
XTRA will show the mA RMS measurement in the display.
MEASURE RF LEAKAGE