Operation Manual
32 GMC-I Messtechnik GmbH
SECULIFEESXTRA
5) Key the DUT. The SECULIFE ESXTRA will show the mA RMS measurement in the display.
Leakage Test 2 - I need to measure the leakage current from the dispersive (REM) port to ground. I need
to have a 200 ohm load from Active to Ground during the test.
Solution:
1) Select Measure RF Leakage from the main menu.
2) If the Leakage selection at the top of the screen does not show Leakage 3, press the button and select
Leakage 3 from the test mode menu.
3) Connect the Active output from the DUT to the Active input on the SECULIFE ES
XTRA. If you are
using the cables from the SECULIFE ES
XTRA accessory kit, use the Yellow Lead.
4) Connect the Dispersive output from the DUT to the Dispersive inputs on the SECULIFE ES
XTRA. If
you are using the cables from the SECULIFE ES
XTRA accessory kit, use the Blue cable without the
clear pin.
5) Connect earth ground port on the SECULIFE ES
XTRA to the ground lug on the back of the DUT. If you
are using the cables from the SECULIFE ES
XTRA accessory kit, use the green cable and green
alligator clip.
NOTE: To see how the SECULIFE ES
XTRA is internally configured, press the Test Setup button. To close
the image, press Test Setup again.