Operation Manual

GMC-I Messtechnik GmbH 33
SECULIFEESXTRA
6) Key the DUT. The SECULIFE ESXTRA will show the mA RMS measurement in the display.
Leakage Test 3 - I need to measure the leakage current from the dispersive (REM) port to ground. I need
to have a 200 ohm load from Active to Ground during the test.
Solution:
1) Select Measure RF Leakage from the main menu.
2) If the Leakage selection at the top of the screen does not show Leakage 3, press the button and select
Leakage 3 from the test mode menu.
3) Connect the Active output from the DUT to the Active input on the SECULIFE ES
XTRA. If you are
using the cables from the SECULIFE ES
XTRA accessory kit, use the Yellow Lead.