Operation Manual

34 GMC-I Messtechnik GmbH
SECULIFEESXTRA
4) Connect the Dispersive output from the DUT to the Dispersive inputs on the SECULIFE ESXTRA. If
you are using the cables from the SECULIFE ES
XTRA accessory kit, use the Blue cable without the
clear pin.
5) Connect earth ground port on the SECULIFE ES
XTRA to the ground lug on the back of the DUT. If you
are using the cables from the SECULIFE ES
XTRA accessory kit, use the green cable and green
alligator clip.
NOTE: To see how the SECULIFE ES
XTRA is internally configured, press the Test Setup button. To
close the image, press Test Setup again.
6) Key the DUT. The SECULIFE ES
XTRA will show the mA RMS measurement in the display.