Operation Manual
GMC-I Messtechnik GmbH 35
SECULIFEESXTRA
Scenario 1 - I need to test the REM function on my generator. The service manual says that the REM
alarm should be Normal at 35 ohms.
Solution:
1) Select REM/ARM/CQM from the main menu.
2) Press the CQM Resistance button and enter 35 in the number pad. The screen should look like the
following:
3) Connect the Dispersive output from the DUT to the Dispersive inputs on the SECULIFE ES
XTRA. If
you are using the cables from the SECULIFE ES
XTRA accessory kit, use the Blue cable with the clear
pin.
4) The DUT REM state should be Normal.
REM / ARM / CQM TEST