Operation Manual

36 GMC-I Messtechnik GmbH
SECULIFEESXTRA
Scenario 2 - I need to test the REM function on my generator. The service manual says that the REM
alarm should occur if the resistance increases by 40%.
Solution:
1) Select REM/ARM/CQM from the main menu.
2) Press the CQM Resistance button and enter a resistance that should result in a normal CQM state, in
this example we will use 50 ohms.
3) Press the Change By Percent button and enter 40 in the number pad. The screen should look like the
following:
4) Connect the Dispersive output from the DUT to the Dispersive inputs on the SECULIFE ES
XTRA. If
you are using the cables from the SECULIFE ES
XTRA accessory kit, use the Blue cable with the clear
pin.
5) The DUT REM state should be Normal.
6) Press the Up arrow under the Change By Percent heading. The resistance will change to 70 ohms and
the DUT should indicate a REM alarm.