Operation Manual

GMC-I Messtechnik GmbH 37
SECULIFEESXTRA
Load Curve Scenario 1 - I need to run a bipolar load curve at 60 Watts with loads from 100 to 1000
ohms, measuring every 100 ohms. I am testing a ForceFx generator and want the SECULIFE ES
XTRA to
automatically configure and trigger the DUT.
Solution:
1) Select Load Curves from the main menu.
2) Select Configure Load Curve from the Load Curve Menu.
3) Press The Setup Loads tab at the top of the screen to configure the Loads
4) Press the Load resistance setting for the First Load and enter 100 in the number pad.
5) Press the Load resistance setting for the Last Load and enter 1000 in the number pad.
6) Press the resistance setting in the Step Loads By option and enter 100 in the number pad.
The screen should now look like the following:
LOAD CURVES