Operation Manual
GMC-I Messtechnik GmbH 39
SECULIFEESXTRA
12)For the DUT setup press the dropdown arrow next to Manual Setup to select Setup By Model.
13)Use the dropdown arrow to set the DUT Output to Bipolar.
14)Use the dropdown arrow to set the DUT mode to the desired bipolar output mode. The screen should
now look like the following:
15)Connect the SECULIFE ES
XTRA to the DUT. Connect the Yellow cable from the Active port on the
SECULIFE ES
XTRA to one of the Bipolar outputs on the DUT. Connect the Green cable from the
Dispersive 1 port of the SECULIFE ES
XTRA to the other Bipolar output on the DUT.
16)Connect the RS-232 cable from the SECULIFE ES
XTRA serial port to the DUT serial port.
17)Connect the Z699B footswitch cable from the DUT to the SECULIFE ES
XTRA.
18)Press Save configuration on the SECULIFE ES
XTRA to save the Load Curve settings to a file for future
use, or Press Start Load Curve to begin the test. The DUT power level and output mode will
automatically be configured through the RS-232 port.