Operation Manual

44 GMC-I Messtechnik GmbH
SECULIFEESXTRA
9) Press the Test Load button again or the Red X button to close the Load Configuration menu.
10)Press the Tolerance option button at the bottom of the screen to configure the tolerance for the RF
measurement.
11)The DUT output mode and output type and power setting can be selected at the top of the screen.
These are informational only and do not have any effect on the test or measurement.
12)Use the dropdown arrow under Expected Reading to change the tolerance mode to Equal to (Value).
13)Press the number button to the right of the expected reading and enter 315 in the number pad.
Additionally, if the desired reading were watts or any other measurement taken by the SECULIFE
ES
XTRA, the red button to the right of the expected reading can be pressed to change the
measurement being tested.
14)Press the number button for High Limit and set the value to 25 with the number pad.
15)Press the number button for Low limit and set the value to 25 with the number pad.
16)Additionally, if you would like to allow the user to override the automated limit testing, you can check
the box next to Allow operator to select Pass Fail status.
17)The screen should look like the following; the configuration is complete for this step. Press Back to
Step Edit and then Done to return to the Autosequence Summary screen.