9) Press the First Load button and enter 120 in the number pad.
10)Press the Last Load button and enter 145 ohms.
11)Press the Step Loads By button and enter 1 in the number pad.
12)Select the option button for Stop when CQM alarm occurs.
13)Press the Set Tolerance button to configure the Pass/Fail limits for the test. Since the requirement is
to alarm at 135 ± 5 ohms, set the CQM Trip point for 135 ohms.
14)Press the Trip Upper limit button and enter 140 in the number pad.
15)Press the Trip Lower limit button and enter 130 ohms in the number pad. The screen should look like
the following:
16)Press the OK button to close the validation requirements menu.
17)The configuration is complete for this step. Press Back to Step Edit and then Done to return to the
Autosequence Summary screen.