Operation Manual

GMC-I Messtechnik GmbH 55
SECULIFEESXTRA
Auto Advance on Pass - This setting allows for automatically stepping through the autosequence if a
measurement passes the tolerance limits that are configured for the selected step.
Auto Capture - When enabled, this setting will automatically capture and validate the DUT output when
the current step is reached during the autosequence. If unchecked, the autosequence will show
instructions to the user and the user must manually select the meter display and capture an RF
measurement.
When SECULIFE ES
XTRA is selected as the trigger source, the activation can be setup manually or by
DUT manufacturer and model. For Manual Setup, the user selects the desired footswitch output to be
used.
If RS-232 is enabled, the user also selects the baud rate and enters commands to be used for DUT setup
and control. When entering commands, use "%P" to represent the power level. In the example below,
the SECULIFE ES
XTRA would send the command ":FTP300" when the autosequence power level is set
to 300 watts.