The user must configure both the SECULIFE ESXTRA meter as well as the measurement tolerance.
To setup the tolerance, select the Tolerance option button at the bottom of the screen.
DUT Output Mode - This setting is for display only, to identify to the user the output mode used.
DUT Output Type - This setting is for display only, to identify the type of output to be measured.
DUT Power Setting - This setting represents what the DUT should be set to for the current step.
Expected Reading - This sets how the SECULIFE ES
XTRA determines whether a measurement passes
or fails. Available settings are Equal To (%), Less Than, Greater Than, or Equal To (Value).
Measurement Units - The user can press the red units button to select what is to be measured for the
current step. Available measurements can be found in the Measure RF Energy chapter, page 72.
High / Low Limit - These settings determine the limits of the measurement. If Expected Reading is set to
"Equal To (%)", the limits will be based on a percentage of the expected reading. If the Expected Reading
is set to "Equal To (Value)", the limits will be based on a measurement offset from the expected reading.
The numbers in parenthesis below the measurement units display the reading limits based on expected
value settings and High/Low Limits.
Allow Operator to Select Pass/Fail - The tolerances configured on this screen will determine the pass or
fail status of the captured measurement during the autosequences. In some cases, it may be useful to
allow the user to select the pass/fail status of a particular measurement. If this box is unchecked, the
operator will not be allowed to modify the pass/fail status of the measurement.