Operation Manual
58 GMC-I Messtechnik GmbH
SECULIFEESXTRA
Load Curve Step - All of the options in the standard load curve are available to the load curve step
configuration.
The tabs across the top of the screen allow the user to configure the Loads, Power settings to be used,
and the Meter configuration for the Load Curve.
Setup Loads Tab - The loads can be configured as either a step based load change or a list based load
change. For the step based loads, the user enters the first load to be used and the last load to be used.
Then the SECULIFE ES
XTRA steps the loads by either a fixed resistance or by a calculated amount to
achieve a fixed number of steps for the test. In the previous picture, the load curve would start at
50 ohms, and increment the load by 35 ohms to provide 20 total steps in the test.
20
50750