Setup Meter Tab - The measurement mode for the load curve can be configured in the Setup Meter Tab.
This screen allows for the selection of the Input Range, Input Mode, and Averaging Rate. These settings
are adjustable to match the waveform or device being tested.
For more information on measurement modes, refer to the RF Measurement chapter of the manual on
page 70.
Manual CQM Step - CQM autosequence steps can be configured as manual or automatic tests. For
manual CQM autosequence steps, the user configures the initial CQM resistance and selects whether the
operator will be allowed to modify the CQM resistance. When the test is running, the operator will need to
determine whether the CQM test passes or fails.