Auto CQM Step - Auto CQM steps allow for greater automation of the autosequence. Auto CQM steps
can be configured as a single resistance test point or as an automatic sweep of CQM resistances where
the CQM status is tested for normal or alarm conditions.
Auto Advance on Pass - This setting allows for automatically stepping through the autosequence if the
CQM status matches the expected setting.
Auto Start CQM Test - When this setting is enabled the SECULIFE ES
XTRA will automatically trigger the
CQM state evaluation to determine if the step passes or fails. If this option is disabled, the operator will
have to trigger the CQM test.
Single Value Tab - The single Value Auto CQM step (shown above) will apply a single resistance to the
DUT CQM circuit and determine whether the step passes or fails based on the expected status.