This screen configures the SECULIFE ESXTRA to measure the RF output of an Electrosurgical Generator.
From this screen the user can configure the measurement, the test load, the parameters on the screen,
and even trigger the DUT.
The measurement mode can be configured using the keys around the top and right side of the screen.
The buttons are described below in a clockwise sequence starting with the back arrow.
Back Arrow - Pressing the back arrow takes the user to the Main Screen.
Hold Key - The Hold key toggles the Hold mode. Measurements are not updated while in the Hold mode.
Measure RF - This button allows the user to select whether the DUT being tested has an isolated output
or ground referenced output. Some generators such as Hyfrecators do not have a return pad and rely on
earth ground as the current return path. When testing these generators, select Ground Referenced
Output and the SECULIFE ES
XTRA will configure the variable load to measure current from the active
input to earth ground.
MEASURE RF ENERGY