Operation Manual
GMC-I Messtechnik GmbH 71
SECULIFEESXTRA
Tissue Test - This button is used to initiate the Tissue Test mode. This mode evaluates the tissue
response function of the DUT. In this mode, the SECULIFE ES
XTRA is configured for an external TRL-
2420 lamp load. The SECULIFE ES
XTRA is connected in series with the load as shown below.
The Tissue Response Test will monitor the current in the load as the lamp heats up. When the current
reaches the Tissue Response Trip point in the DUT, the generator output will automatically deactivate. At
this point, the display will change to show the trip current as well as a graph of the current during the test.
See page 18 for typical connections.
Test Setup - This button shows the connection diagram for the RF Measurement mode.
Display Parameters - This key allows the user to select the number of measurements, or zones, that are
shown on the screen. 6 Screen configurations are available, 5 Display Screens which have 1, 2, 3, 4 and
5 display zones respectively, and a Measurement List Screen which shows all available measurements.
Each Display Zone can be customized to show the desired parameter from the following options: