Mode Parameter Abbreviation Description
Pulsed and Continuous Mode
mV RMS mV
This is the mV measured directly from the RF current
transformer.
mA RMS mA
This is the converted mA measurement based on the
RF transformer mV to mA attenuation ratio.
Power in Watts Watts
This is the computed power based on load setting and
mA measured.
mV Peak mV Pk
This is the maximum mV measured in the buffer.
NOTE: This is shown as an absolute value.
mV Peak - to - Peak mV P-P
This is the difference between maximum mV
measured and min mV measured.
mV Peak /
Peak - to - Peak
Pk/P-P
This is the ratio of Peak versus
Peak to Peak millivolts
mV
Positive Peak Only
mV Pk+
This is the maximum positive mV measured in the
buffer. For asymmetric waveforms this can determine
if the output polarity is reversed.
Crest Factor CF
This is the ratio of peak to rms of the measured
waveform.
Pulsed Mode Only
Time - Pulse On
T
on
This is the duration that the pulsed waveform is on
(see Diagram 1)
Time - Pulse Off
T
OFF
This is the duration that the pulsed waveform is off
(see Diagram 1)
Time - Total Cycle
T
cyc
This is the total cycle time of the pulsed waveform
(i.e. T
on
+ T
off
) (see Diagram 1)
% Duty Cycle %Duty
This is the ratio of the pulse on time (T
on
) versus cycle
time (T
cyc
) (see Diagram 1)
mV Pulse
mV
cyc
This represents the RMS mV over one pulsed cycle
(see Diagram 1)
mA Pulse
mA
cyc
This represents the RMS mA over one pulsed cycle
(see Diagram 1)
Watts Pulse
W
cyc
This represents the RMS Watts over one pulsed cycle
(see Diagram 1)
Ton
Toff
Tcyc
mV
mA
Watts
CF
mVcyc
mAcyc
Wcyc
Diagram 1 (Pulsed mode only)