Operation Manual
74 GMC-I Messtechnik GmbH
SECULIFEESXTRA
Averaging - This key allows the user to select the FAST, MEDIUM, or SLOW averaging mode. Fast
averaging will provide a quick response to incoming signals. Slow averaging will provide a more stable
display, but will be slower to respond to small changes in the RF input signal. The value for the Averaging
settings can be configured in the System Setup screen.
Input Range - Allows the user to select the range for the RF input. The range can be selected as 100mV,
1000mV, or Auto range.
Input Mode - Allows the user to select the mode for the RF measurement. The mode can be selected as
Continuous or Pulsed. Continuous RF waveforms include Cut, Bipolar, or Coag modes. Pulsed RF
waveforms consist of a long period duty cycle where the RF output might be active for 20 mS and inactive
for 500 mS.
Test Load - This button shows a menu that allows the user to select internal load mode, external load
mode, or the combination of internal/external loads. The menu also allows for the selection of the internal
and external load values.