Operation Manual
GMC-I Messtechnik GmbH 75
SECULIFEESXTRA
Generator Type - The buttons labeled "Isolated" and "Gnd Ref" allow the user to select whether the DUT
being tested has an isolated output or ground referenced output. Some generators such as Hyfrecators
do not have a return pad and rely on earth ground as the current return path. When testing these
generators, select Gnd Ref and the SECULIFE ES
XTRA will configure the variable load to measure
current from the active input to earth ground.
Configure Footswitch - This button opens the footswitch configuration menu that will select the
footswitch output and duration that is used when the footswitch trigger button is pressed. To close the
footswitch setup menu, press the Configure Footswitch button.
Activate Footswitch - This button is used to trigger the footswitch output for the duration that is
configured in the Footswitch Setup menu. The footswitch can be deactivated before the activation timeout
period by pressing the Activate Footswitch button a second time. While the footswitch output is active, the
Activate Footswitch button will remain highlighted with a yellow background.