The power load curve allows the user to test and graph the power output of the DUT. The loads used and
DUT power settings are programmable. The SECULIFE ES
XTRA can be configured to trigger the DUT
using one of the five footswitch relay outputs. For further automation, the RS-232 port can be used to
configure the DUT output mode and power level during the Load Curve.
The load curve menu allows the user to configure a load curve, run the factory default load curve, run a
saved load curve profile, or view a load curve that was saved previously.
Load Curve Configuration
The tabs across the top of the screen allow the user to configure the Loads, Power, Meter and ESU
generator.
POWER LOAD CURVES