Load Config Tab – The loads can be configured as either a step based load change or a list based load
change. For the step based loads, the user enters the first load to be used and the last load to be used.
Then the SECULIFE ES
XTRA steps the loads by either a fixed resistance or by a calculated amount to
achieve a fixed number of steps for the test. In the picture above, the load curve would start at 50 ohms,
and increment the load by 35 ohms to provide 20 total steps in the test.
For list based loads, the following screen is shown, allowing the user to select specific resistances to be
used for the test.
Single Power Level Tab - For the load curve power configuration, the user can set the SECULIFE
ES
XTRA to run the load curve at a single power level or multiple power levels. The following screen shows
the configuration for a single power level.
By pressing on the power setting, the user will be able to enter the desired power level for the load curve
test.